Brand: Hanna
Made in – China
Range pH 0.0 to 14.0 pH
resolution pH 0.1 pH
accuracy pH ±0.1 pH
Remove the protective cap. Do not be alarmed if white crystals appear around the cap. This is normal with pH electrodes and they
dissolve when rinsed with water.
Turn the meter on sliding the switch on the top.
Immerse into solution up to the max immersion level.
Stir gently and wait until the display stabilizes.
After use, switch off pH meter, use the water clean the electrode and replace the protective cap.
DO NOT USE DISTILLED OR DEIONIZED WATER FOR STORAGE PURPOSES.
Large fluctuations in readings could be due to lack of calibration, dry electrode or rundown batteries.